How do you validate a micropipette?

How do you validate a micropipette?

The most common way to check your pipette accuracy is by weighing water. The density of water is 1 g/mL. This means that every microliter (µL) should weigh 0.001 g. In other words, if your pipette is accurate, the amount of water you dispense will equal the amount the water weighs.

What are the methods of pipette calibration?

The calibration of pipette is carried out by gravimetric method. When determining the volume of water, the accuracy of measurements is effected by ambient temperature, atmospheric pressure and relative humidity. These factors are usually combined to give the Z factor, used in calculation of volume of water.

What is the tolerance of a micropipette?

The simplest and most reliable solution is to replace the 200 µL pipette with a 100 µL pipette. This pipette, when used at its full scale setting, can be tested against a 2% tolerance.

How do you calculate a pipette?

Use the formula V = w * Z to calculate the volume dispensed by the pipette. W is the weight of the distilled water. Z is the “conversion factor based on the density of the water” and V is the calculated volume of dispensed water.

How do I make my pipette reading more accurate?

How good are you?

  1. Pre-wet the pipette tip.
  2. Immerse the tip to the proper depth during aspiration.
  3. Pause consistently after aspiration.
  4. Use consistent plunger pressure and speed.
  5. Pull the pipette straight out.
  6. Examine the tip BEFORE dispensing a sample.
  7. Examine the tip AFTER dispensing a sample.
  8. Use standard mode pipetting.

Are pipettes accurate?

A pipette is accurate to the degree that the volume delivered is equal to the specified volume. A pipette can be consistently inaccurate but this inaccuracy could be very precise, for example if a pipette reads consistently low.

How do you calibrate a plastic pipette?

A small plastic or glass dropper pipet can be calibrated by counting the number of drops of water it takes to make a certain volume.

  1. Fill the pipet.
  2. Counting each drop, squeeze the pipet gently to release water drop by drop into a dry 10 mL graduated cylinder.

What is a pipette tolerance?

A volumetric pipette, bulb pipette, or belly pipette allows extremely accurate measurement (to four significant figures) of the volume of a solution. The tolerance depends on the size: a 0.5-mL pipette has a tolerance of ±0.006 mL, while a 50-mL pipette has a tolerance of ±0.05 mL.

How do you find the tolerance of a pipette?

Relative error tolerance limits at other volume settings can be calculated by dividing the absolute tolerance limit (see Table 1 or Note b) by the set point volume. Multiply the result by 100 to convert it to a percentage.

How do you calculate the accuracy of a pipette?

Calculate the accuracy of the pipette. To calculate the accuracy you can use the equation A = 100 x Vavg/V0, where A is the accuracy of the pipette, Vavg is the average calculated volume and V0 is the value you set the pipette to dispense. Accuracy should be between 99-101%.

What are micropipettes used for?

Micropipettes are a common type of laboratory equipment designed to transfer and measure very small amounts of liquid. They can be used to measure volumes as low as 0.003 ounce (0.1 milliliter). Standard micropipettes used in laboratory settings feature a few basic parts including an adjustment dial and a plunger button.

How do you adjust the volume on a micropipette?

Volume adjustment on Micropipette. For adjusting the proper volume rotate the volume adjustment dial present at the top of the micropipette. It can be rotated right or left to decrease or increase the volume. The volume that is going to be measured by micropipette is shown on the digital layout.

How do I choose a micropipette for my experiment?

You can either use a micropipette whose volume is already fixed or one whose volume can be adjusted as per requirement. If you regularly use a single volume such as 100 µl for your experiments, go for a fixed volume pipette like MICROLIT RBO Fixed Volume (Single Channel).

What is the difference between micro and medium white micropipette tips?

Micro white tips are used for P2 and P10 size micropipette used for taking volume from 0.2 µ to 10 µ. Medium white micropipette tips are used for P20 micropipette measuring a volume range of 2 µ to 20 µ