What does EE mean on a blood pressure monitor?
What does EE mean on a blood pressure monitor?
over-inflated
The EE symbol with a down arrow indicates the cuff is over-inflated. In most instances the cuff has been wrapped too tight or you are using the incorrect cuff size. Turn the monitor off. Read the instructions to correctly apply the cuff. Wait five to ten minutes then take another measurement.
Why does my blood pressure monitor keep saying error?
The most common error when using indirect blood pressure measuring equipment is using an incorrectly sized cuff. A BP cuff that is too large will give falsely low readings, while an overly small cuff will provide readings that are falsely high.
What does E1 error mean on Omron Blood Pressure Monitor?
Device shows ‘E1’ Air plug is not connected or air is leaking from the arm cuff. Check air plug connection. Take another measurement and remain still until the measurement is complete.
What does Error 5 mean on Omron Blood Pressure Monitor?
clothing is interfering with
Per the book, E5 means that clothing is interfering with the cuff. The error code did not occur when I repeated the measurement without doing anything to the cuff.
What does E4 error mean on Omron Blood Pressure Monitor?
For Error E4 it says: “Hand shaking while measurement; Hands or body shaking during measurement.” Under possible solutions it says, “keep static position and try again.”
What does E4 mean on Omron?
Movement during measurement
Omron blood pressure monitor error codes
Error code / symbol | Why it happens |
---|---|
E3 | The arm cuff inflated past than the maximum pressure allowed and deflated automatically (applies to manual cuff inflation) |
E4 | Movement during measurement |
E5 | Clothing is disrupting the measurement |
Er | Device error |
What does Error 4 mean on blood pressure machine?
Er 4. Air leak in the pneumatic system or the cuff is not tight. Attach the cuff correctly and try again.
What is e3 on BP machine?
The cuff is not properly secured. Refasten the cuff, wait 3 minutes and then measure again.