What does cycle fault 16 mean on a Statim 2000?
What does cycle fault 16 mean on a Statim 2000?
Cycle Fault 16 being displayed on your Statim indicates the steam generator (boiler) temperature went above 171°C during the heat up phase of a cycle. The most common reason for this fault is caused by not enough water entering the boiler from the water pump.
How do I calibrate my Statim 2000 boiler?
How To Recalibrate The Statim 2000 Boiler
- Turn OFF main power and remove the cover from Your Statim.
- Short together PINs W1 with a “jumper” on the main PCB board.
- Turn ON the power.
- The LCD should display CALIBRATION with: 2 digits in the upper left, 2 digits in the upper right.
- Insert The Cassette.
How do you drain a Statim 2000?
To drain the water from the reservoir of the Statim 2000, use a high volume evacuation (HVE) hose from a vacuum system. Insert the hose into the reservoir and allow it to suck the water out. If a vacuum system is not available the water moat he siphoned from the reservoir.
How does a Statim sterilizer work?
STATIM provides sterilization and dryness at speeds faster than conventional chambered autoclaves. Using only a small amount of water for each cycle, the STATIM heats and converts to steam in mere seconds. The innovative design of STATIM’s fully removable chamber allows for quick heating and cooling.
How do you clean a Statim sterilizer?
Proper Care of the STATIM Cassette
- Clean the cassette once a week with mild, chlorine-free dishwashing soap.
- Spray the cassette with Stat-dri (or equivalent) after each cycle to promote faster cassette drying.
- Lubricate the rubber cassette seal once or twice a week with a mild liquid soap—not antibacterial soap.
What kind of sterilization is a STATIM?
steam flush pressure pulse
The STATIM is a steam flush pressure pulse (SFPP) Sterilizer. With this type of Health Canada licensed technology, air is actively removed from the chamber through pressurizations and flushes with steam. No vacuum is drawn and the chamber pressure does not drop below atmospheric.
How does a STATIM sterilizer work?
What does the code 16 mean on a Statim 2000?
Statim Cycle Fault 16 – Statim 2000 Error Code Cycle Fault 16 “What It Means And How To Fix It” Cycle Fault 16 being displayed on your Statim indicates the steam generator (boiler) temperature went above 171°C during the heat up phase of a cycle. This is caused by
What does cycle Fault 16 mean on a Statim 2000?
Statim 2000 & 5000 Error CodeCycle Fault 16″What It Means And How To Fix It”. Cycle Fault 16 being displayed on your Statim indicates the steam generator (boiler) temperature went above 171°C during the heat up phase of a cycle. The most common reason for this fault is caused by not enough water entering the boiler from the water pump.
How do I read the error codes on my statim sterilizer?
There are 2 different sets of error codes: Original Error Codes are prefaced by “Check Cassette” followed by a number &s New Error Codes are prefaced by “Cycle Fault” followed by a number. The Statim Sterilizer should never leak steam or have water dripping under the machine.
How do I test my statim 2000?
Lift up on the rear of the housing and slide it forward gently. Once cleared, you can pick it up and set it on it’s side (be careful, there are ribbon cables connected..this is where the Statim 2000 Extension Test Cable comes in handy Start with the end of the pump with the larger (gray) tubing connecting to the reservoir.